Revised 18 Feb 2010 by A. R. Walker
| Contents: | |||
| 1. | Weather and thermal issues | ||
| 1.1 | General closure conditions | ||
| 1.2 | Dome opening | ||
| 1.3 | Lateral sliding doors opening | ||
| 1.4 | Primary mirror cooling | ||
| 1.5 | Floor cooling ans stirring fan on main floor (M) | ||
| 1.6 | Oil cooler | ||
| 1.7 | Chimney fans | ||
| 1.8 | Air conditioning | ||
| 1.9 | Temo4m and weat | ||
| 2. | Brief reminder of known optical problems | ||
| 2.1 | Astigmatism | ||
| 2.2 | Coma | ||
| 2.3 | Defocus | ||
| 3. | Prime Focus nights | ||
| 4. | F/8 nights | ||
ACTR#002/00
DT: 18th October 1999
TO: Telops, Scistaff, ETS leaders
FM: ACTR and Maxime Boccas
RE: Revised operation instructions at the 4m-telescope for image quality program
Since the end of 1993, many improvements have been carried out at the 4m telescope to improve its image quality. The principal changes are: an active primary mirror support with lookup tables, a repolished f/8 secondary mirror, an image analyzer for the Cassegrain foci, removal of most heat sources inside the dome, active control of the primary temperature during the day, active ventilation of the primary cell at night, dome ventilation doors and active control of oil temperature.
The night assistant must check that all the environmental control functions explained below are working properly during the night.
ABSOLUTE closure when:
Dome re-opening:
In moderate wind: if wind > 25 mph, avoid observing in the wind direction as it could jitter the telescope and will bring dirt inside the dome.
BEWARE PLEASE: Don't permit observing beyond these limits! If you have trouble to be heard by the astronomer, please REPORT it (so we can politely explain the rules).
The dome AND the mirror cover have to be opened 30 minutes before sunset, orienting the shutter eastward to make sure no direct sunlight will enter the dome. Dome and mirror cover have to be closed at the end of the night.
These doors have to be opened just after sunset and left opened all night.
During the night, you are invited to type "temp4m" and check that:
1/ Tmirror < Tgoal - 2.5° : glycol valve closed (but ambient air is still blown onto the mirror)
2/ Tgoal + 1.7° < Tmirror < Tgoal -2.5°: glycol valve partially opened
3/ Tmirror > Tgoal + 1.7 ° : glycol valve fully opened
4/ If Tmirror < Tdew_point + 7 ° : the glycol valve will close (safe limit to prevent condensation)
The cooling of the C floor and M floor in the dome is always ON. It is currently regulated manually.
The large stirring fan on the M floor (inside the dome) should always be ON when the dome is closed and OFF when the dome is opened. This fan improves greatly the air circulation inside the dome. It can be turned off during instrument change or any other works around the telescope if the noise bothers you. But turn it back on when you are done!
The cooling of the oil circulating under the R.A. rear and front pads is activated automatically when the pumps are turned ON by Observer support at sunset. The oil glycol valve is controlled (range is 1-2.25 V) by an equation involving the low dome and oil glycol temperature sensors. It is important that the oil cooling be working: if it is not, you can diagnostic it in "temp4m" (the "before pad" and "glycol" entries on the OIL line will quickly indicate around 30°).
These fans are located inside the chimney at the level of the primary mirror and suck ambient air into the mirror cell so as to form a laminar downward flow in the chimney. This is to prevent Cass cage heat sources from creating convection in the light path in the chimney.
At Cassegrain foci (f/8), Observer Support has to turn ON the chimney fans before sunset. The switch is on the power supply on the old console room desk. The night assistant must turn it OFF at the end of the night.
The current thermal plan in the building is not changed. Always leave the air conditioning at full power. Do not use heat sources at level M, MZ and C. Always maintain the doors closed in the passageways (especially the glass door in front of the lift at GR floor, the cryocooler compressor and pump rooms). In general, respect the signs in the building.
Until we have a graphical web interface, type "temp4m" and "weat" to check the temperatures and weather conditions. Type, "/ut02/seeing/temper/tlog4m" for graphical analysis (the script should work at least on ctiot0 and ctioa2). We are working on a web page for thermal related issues.
The image quality of the telescope can suffer deterioration mainly from 3 optical problems:
This is due to tiny slippage of M1 and/or deformation under its own gravity. Astigmatism will show up as elongated images, perpendicular on each side of the focus, especially for large H.A. (typically more than 2 hours). We compensate that with a lookup table, which controls the pressure in the air bags under M1. Thus the TCP toggle "CORR" should always be ON, at all foci. Note that if the F/8 secondary mirror lose its vacuum, strong astigmatism will immediately show up (typically 2-4 microns as measured by iman).
This aberration shows up in 2 cases: at Prime Focus if the primary mirror axis is not aligned with the PFADC and in the 2-mirror configuration (f/8) when there is a misalignment between them. Coma produces images with a flare (a tail like a comet) which is identical (same amplitude and direction) on each side of the focus.
At Prime Focus, we don't have a "quick-fix" tool to correct coma so any problem need to be reported (it hasn't happened, though).
Coma usually increases the further you go from the optical axis -this is called "field coma" and is normal (it is always supposed to be 0 on axis)- and also shows up when the optical axis of M1 and M2 are not coaligned -this is called "decentering coma" and is abnormal-. Our f/8 is a Ritchey-Chretien (RC) type, which means it is optically designed to correct the field coma. If the telescope is properly aligned there should not be any visible coma at f/8. If coma shows up anywhere in the field (it would usually be constant all over the field), there is a mirror misalignment and this can be corrected by using IMAN.
IMAN detailed instructions are on the web at Instructions for normal use and in the written manual next to the TCP ("User documentation: active optics system").
Thermal expansion that can occur at night will cause the telescope to defocus. The prime focus changes at a rate of -110 microns/°C, the f/8 focus at -780microns/°C. The actual movement of He f/8 mirror is a factor 10 less (i.e. -78 microns/°C)
Use "temp4m" to check the Serrurier truss temperature and adjust the focus accordingly.
Mosaic can automatically check the Serrurier truss temperature and adjust the telescope focus accordingly before each exposure.
* CORR ON all the time.
* Temperature drift compensation:
Focus number increases when focal plane moves away from M1
Focus number is in microns neglecting the effect of the PFADC (the PFADC has a power of 1.076, i.e. multiply the focus change by 1.076 to get the real motion in microns in the focal plane).
Decrease focus setting by 110 units per 1° increase in temperature
* CORR ON all the time.
* The toggle next to f/8 on the TCP status display should be OFF all the time (this is for an eventual coma lookup table ; if ever you need to turn it off, go in the TCP menu, and select /Active_Control/Tilt_Secondary/On-Off_Auto_Tilt/Off).
* Temperature drift compensation:
Focus number increases when focal plane (and M2) moves up
The focus number represents microns of motion of the focal plane (not of M2)
Decrease focus setting 780 units per 1° increase in temperature
*Collimation:this has to be checked according to the following procedure:
The images are automatically stored in the appropriate archive.
One image should be taken at the beginning of the night, after doing the telescope zpoint and focussing the instrument. ). The spectro mode of Osiris needs to be focused in the imaging mode (both modes are parfocal), thus it is easy to take that first seeing sequence.
A second image should be taken by the time it is 1 hour before morning twilight, preferably after the focus has been rechecked. If Osiris is used in spectro mode, you obviously need to change back to imaging mode to take this second seeing frame.
In case some problems forced you to start late (i.e. after 11PM), only a single measurement will be required.
For the time being, no seeing measurement is required with IRS. In the near future, we hope to do the seeing measurement with the Tip-Tilt camera for all IR instruments.