The 1.5-m Cass Spectrograph with Loral 1200 x 800 CCD


Commissioning of the Loral 1200 x 800 CCD on the 1.5-m Cass Spectrograph ========================================================================

Final commissioning of the Loral 1200 x 800 CCD (hereinafter Loral 1K CCD) plus 1.5-m spectrograph combination was carried out during two engineering nights on November 15-16, 1995. A previous engineering run in May (reported on pages 25-26 of NOAO Newsletter No. 43, see below) had allowed an initial evaluation of the image quality, throughput, fringing characteristics and flexure of the system at the telescope. The encouraging results of that engineering night were amply confirmed on the two commissioning nights in November -- the performance of the Loral 1K CCD is superior in all aspects to that of the front-illuminated GEC CCD used for so many years with the 1.5-m spectrograph, offering 40% greater wavelength coverage (at the same resolution) and >= 2-3 times the quantum efficiency.

Within the next few months we will attempt to have a new user manual for the 1.5-m spectrograph + Loral 1K CCD available on the CTIO WWW pages. Twelve gratings are available which give the following wavelength resolution and coverage with the Loral 1K:

                                  FWHM     Wavelength
     Gtg     l/mm    blaze $   Resolution   Coverage
   --------------------------------------------------
     13      150      5000       17.2 Å      6900 Å
     11 *    158      8000       16.4        6550
     09      300      4000        8.6        3450
     32      300      6750        8.6        3450
     22 *    300     10000        8.6        3450
     58      400      8000        6.5        2590
     16      527      5500        4.8        1965
     26      600      4000        4.3        1725
     35      600      6750        4.3        1725
     56      600     11000        4.3        1725
     47      831      8000        3.1        1245
     36 &   1200      7500        2.2         860    

     $ Blaze is first order Littrow blaze. Effective blaze 
       wavelength when used in the 1.5-m spectrograph is 0.89 
       of the Littrow value.

     * silver coated does not reflect light below ~4000Å 

     & Cannot be tilted far enough to be used in II order

The Loral CCD has 15 micron pixels and a slit width of 143 microns (2.6 arcsec) projects to 2 pixels. In practice, however, the focus of arc lines is observed to range between 2-3 pixels (FWHM) due to the camera optics and charge diffusion within the CCD (for more details, see the article in Newsletter No. 43). The wavelength resolution figures in the above table are calculated for a FWHM of 3 pixels.

Efficiency measurements of the total telescope/spectrograph/CCD combination were obtained with gratings 09 and 32 during the November engineering nights. The results are given in the following table in terms of the percentage of photons striking the telescope primary mirror which are eventually detected by the CCD. Note that these numbers are still not definitive due to uncertainties in the precise value of the CCD gain on the engineering night; however they should be accurate in a relative sense. We hope to be able to provide final numbers within the next 6 months.


                               Grating
        Wavelength           09      32
       -----------------------------------
          3500               7.5 %      
          4000              13.5        
          4500              16.5        
          5000              17.0         
          5500              16.0         
          6000              13.5         
          6500              10.0    14.8 %
          7000                      13.6  
	  7500                      11.8
	  8000                       8.5
	  8500                       6.5
	  9000                       5.0
	  9500                       2.9
The upper right amplifier does not perform satisfactorily, therefore the CDD is read in single-channel mode (through the lower left amplifier). As part of the commissioning process, a reduction of 10 microseconds per pixel was achieved. The resulting gains, read noises, and readout times are:
       1/Gain    Read Noise   Read Time
      (e-/ADU)     (e-)       (seconds)
     -----------------------------------
        4.11       7.71         15.6
        2.87       7.11         19.7
        2.05       6.50         25.8
        1.42       6.13         34.0
        0.96       5.88         46.3
Full well of the Loral 1K CCD is 118,000 e-. Over this range, the CCD delivers excellent linearity (gain variation = 0.26% peak-to-peak).

As reported in Newsletter No. 43 (see below), the Loral 1K CCD fringes with substantial amplitude at wavelengths redward of 7500Å. However, thanks to the lack of significant flexure in the 1.5-m spectrograph and camera, it is possible to remove nearly all of the fringing using normal dome flats.

The accompanying figure (above) shows spectra of two type II supernovae obtained with the 1.5-m + Loral 1K commbination. These spectra are the sum of separate blue and red observations obtained at ~8.5 Å resolution with gratings 09 and 32. Total integration time for SN 1995ad was 30 minutes in the blue and 30 minutes in the red, while SN 1995V was observed for 90 minutes in the blue and 45 minutes in the red. Approximate magnitudes measured from the spectra are B = 16.8 and V = 15.7 for SN 1995ad and B = 18.7 and V = 17.6 for SN 1995W. Note the lack of obvious residual fringing at red wavelengths. In the case of SN 1995ad, a signal-to-noise value of 35:1 was obtained at a wavelength of 9250Å where the CCD fringing amplitude reaches 20%. (The signal-to-noise in the these spectra is limited more by photon statistics than any residual fringing.) With the successful commissioning of the Loral 1K CCD which is controlled by an Arcon, the last VEB controller in service on Tololo has now been officially retired! Que en paz descansen...

-- mphillips@noao.edu, sheathcote@noao.edu, rsmith@noao.edu


NOTE: Preliminary Information: the following is based on the results of the first engineering night with this system plus laboratory testing.


Initial tests of the Loral 1200 x 800 CCD (hereinafter Loral 1K CCD) plus 1.5-m spectrograph combination, were made during an engineering night on May 17. The science grade chip had only been installed a few days prior to this engineering run so characterisation of its properties was far from complete. However, its performance was judged sufficient to permit an initial evaluation of the image quality, throughput, fringing characteristics and flexure of the system at the telescope. The engineering run was highly sucessful and the measured performance is very encouraging. We expect that the Loral 1K CCD will be made available for visitor use following a second engineering run early in the second Semester of 1995. Thereafter it will be the only CCD offered with the 1.5-m spectrograph. To help in the planning of observations scheduled during second Semester the following discusses what is currently known about the performance of the Loral 1K CCD and compares it to that of the present GEC CCD.

Gratings, Resolution & Coverage:


Gtg	l/mm	blaze	     GEC	    Loral
			Å/pix 	Cover	Å/pix	Cover

13	150	 5000	8.4	4800	5.73	6820
11 *	158	 8000	8.0	4600	5.45	6530
09	300	 4000	4.21	2400	2.87	3410
32	300	 6750	4.21	2400	2.87	3410
22 *	300	10000	4.21	2400	2.87	3410
58	400	 8000	3.15	1800	2.15	2560
16	527	 5500	2.36	1350	1.61	1920
26	600	 4000	2.11	1200	1.44	1705
35	600	 6750	2.11	1200	1.44	1705
56	600	11000	2.11	1200	1.44	1705
47	831	 8000	1.5	 860	1.02	1220
36 &   1200	 7500	1.05	 600	0.72	 850	

% Blaze is first order Littrow blaze. Effective blaze 
  wavelength when used in the 1.5-m spectrograph is 0.89 
  of the Littrow value.

* silver coated does not reflect light below ~4000Å 
& Cannot be tilted far enough to be used in II order

The GEC CCD has 22 micron pixels; a slit width of 210 microns (3.8 arcsec) projects to 2 pixels. With this device there is no evidence that the resolution of the spectrograph is limited by either the camera optics or the MTF of the detector. The measured FWHM of comparison lines corresponds very closely to the projected width of the spectrograph slit down to the Nyquist sampling limit, and 2 pix FWHM resolution is routinely achieved. There is little variation of image quality with position on the chip, or with wavelength.

The Loral CCD has 15 micron, pixels and is 1.4 times longer than the GEC; a slit width of 143 microns (2.6 arcsec) projects to 2 pixels. Because of the finer sampling and larger size of this CCD it is expected that the camera optics will somewhat limit the resolution, especially at the extreme edges of the field. In addition at KPNO they have been unable to get images better than ~3 pix FWHM with their Loral chips. This has been attributed to diffusion of photoelectrons within the CCD. This effect is greatest at blue wavelengths since higher energy photons are absorbed closer to the surface of the CCD.

The following table shows the measured FWHM of arc lines obtained for a single tilt of grating 32 for a slit width 110.5 microns (2 arcsec) showing the dependance on position on the chip (and wavelength). Values are given for the center of the slit (Y=200) and at the two extreme edges (Y=130, 270).

FWHM (pix) as a function of position
====================================
Line	  X		  Y
(Å)	 (pix)		(pix)
		130	200	270	
====================================
4471	 112 |  3.12    2.78 	2.82
4764	 215 |  2.96    2.86 	2.70
5015	 303 |  2.89    2.69 	2.70
5876	 602 |  2.65    2.42 	2.46
6678	 879 |  2.52    2.13 	2.25
6965	 978 |  2.47    2.04 	2.47
7384	1121 |  2.62    2.13 	2.46
7635	1207 |  2.97    2.13 	3.14
====================================
In general, although the lines are wider than the projected slit width, and there is some variation with position, the resolution with a slit width of 2-3 arcsec is better than or comparable to what would be obtained with the GEC CCD and the same grating.

The graph shows the FWHM as a function of slit width for spectral lines at the center of the CCD (in X and Y). Curves are shown for 3 wavelengths 3888Å, 6678Å and 9224Å

QE:

        GEC     Loral 1K
3000    20      25   
3500    19      48   
4000    17      65   
5000    22      83   
6000    35      93   
7000    45      91   
8000    30      83   
9000    14      59   
10000    3      10   

System efficiency:

The following are the measured system efficiencies (percentage of photons striking the telescope primary mirror which are eventually detected by the CCD) for the 1.5-m spectrograph with the GEC CCD using gratings 11 and 13


	   GEC			 Loral 1K
        11      13		 11	 13
3000    0.0     1.1		 0.0	 1.4
3500    0.2     1.6		 0.5	 4.0
4000    0.9     2.0		 3.4	 7.6
5000    2.2     2.7		 8.3	10.2
6000    7.0     4.9		18.6	13.0
7000    6.4     3.6		12.9	 7.3
8000    3.6     2.0		 9.9	 5.5
9000    1.5     0.7		 6.3	 2.9
10000   0.4     0.0		 1.3	 0.0

The numbers for the Loral were estimated by scaling by the ratio of the QE's given above.

Unfortunately the engineering night had heavy cirrus / thick cloud. Therefore we do not have any measurements of the absolute sensitivity. However, observations of standard stars confirm that the sensitivity peaks at approximately 6000Å, where the QE curve peaks, and that there is significant sensitivity down to the atmospheric cutoff at 3000Å.

RON & Dark Current:

So far only the lower left amplifier has been comissioned and the CCD is being read in single channel mode.

The following table shows the gain (e-/ADU), RON (e-), and readout time (s) for the currently available gain settings.

                             Arcon3.9 == Loral 1K (1200*800)
       i
       n                                                  Bin1x1  Maximum
       d  DCS  Delay  Read_Noise    1/Gain -> Read_Noise   Read   Linear
       e time            (ADU)      (e-/ADU)     (e-)      Time   Signal 
       x (us)          LL    UR     LL   UR    LL     UR    (s)   (ADU)
         ----  -----  ----------   ---------  ----------   ----   -----
  1:   1   5     3     2.5   0     4.11   0   10.3    0    25.1   21900
  2:   2   7     3     2.6   0     2.87   0    7.6    0    29.1   31400
  3:   3  10     3     3.6   0     2.05   0    7.2    0    35.3   43900
  4:   4  14     3     4.9   0     1.42   0    6.9    0    43.4   63400
  5:   5  20     3     6.7   0     0.96        6.5    0    55.6   65534
Full well (~90,000 e- is reached before the ADC saturates at the higher gains (more e-/ADU). The Non-linearity (peak-to-peak gain variation) is believed to be less than 2% for levels below full well / ADC saturation.

Currently the dark current is very high ~15e-/pixel/hour. However, it is expected that this will be reduced to a few e-/pixel/hour by opperating the CCD in MPP mode and by running it at a lower temperature. Fringing:

The Loral 1K CCD fringes with substantial amplitude at wavelengths redward of 7500Å. Press here if you realy want to be horrified . The fringes run approximately perpendicular to the dispersion. The peak-to-peak amplitude and fringe spacing are given in the following table and shown in the accompanying graph:

Wavelength	Amplitude	Spacing
(Å)		(%)		(Å)
=======================================
7500		 2.5		38
7750		 2.8		40
8000		 4.5		40
8250		 8.4		46
8500		11.3		38
8750		14.6		42
9000		16.0		54
9300		20.6		54
9500		19.6		56
9750		17.0		66
10000		11.4		50
10500		 7.4		60
=======================================
At least in flat field frames the fringe amplitude does not depend on slit width. Nor does it depend (to first order) on the position where the light of a particular wavelength falls.

We do not yet know how well fringing can be corrected by flat fielding techniques. Given the above amplitudes it seams likely that for wavelengths shortward of about 8000Å fringing is unimportant or easily correctable. Redward of this it is likely that it will be necessary to obtain quartz flats for each object and take great care in flatfielding the data. Even then, observations requiring high S/N at wavelengths redward of 8000Å may not be possible with this CCD.

Flexure:

The following table shows the displacements (pixels) due to flexure parallel and perpendicular to the dispersion as a function of Hour Angle and Declination.

			        Parallel to Dispersion

===============================================================================
      |                               HOUR ANGLE
 DEC  |              WEST                                     EAST
===============================================================================
 +30  |                                   0.0h
      |
      |                                  +0.10
------|------------------------------------------------------------------------
   0  |               3.0h                0.0h                3.0h
      |
      |              +0.35               -0.16               -0.23
------|------------------------------------------------------------------------
 -30  |     4.5h      3.0h      1.5h      0.0h      1.5h      3.0h      4.5h
      |
      |    +0.75     +0.42     +0.14      0.00     -0.09     -0.10     -0.06
------|------------------------------------------------------------------------
 -60  |   5.0h                            0.0h                            5.0h
      |
      |  +0.79                           +0.13                           -0.06
------|------------------------------------------------------------------------
 -90  |                                   0.0h
      |
      |                                  +0.68
===============================================================================


			      Perpendicular to Dispersion

===============================================================================
      |                               HOUR ANGLE
 DEC  |              WEST                                     EAST
===============================================================================
 +30  |                                   0.0h
      |
      |                                  -0.02
------|------------------------------------------------------------------------
   0  |               3.0h                0.0h                3.0h
      |
      |              -0.66               +0.08               +0.34
------|------------------------------------------------------------------------
 -30  |     4.5h      3.0h      1.5h      0.0h      1.5h      3.0h      4.5h
      |
      |    -0.75     -0.61     -0.29      0.00     +0.26     +0.34     +0.37
------|------------------------------------------------------------------------
 -60  |   5.0h                            0.0h                            5.0h
      |
      |  -0.81                           -0.10                           +0.40
------|------------------------------------------------------------------------
 -90  |                                   0.0h
      |
      |                                  -0.49
===============================================================================

sheathcote@noao.edu mphillips@noao.edu rschommer@noao.edu rsmith@noao.edu